114 Appendix
Appendix B: Adding a code resistor to
a custom sensor
In order for TriLog to automatically identify a sensor, it must have a code resistor.
Figure 38 shows the TriLog sensor plug configuration.
You must connect the resistor between the ID resistor input and the GND input.
Figure 38: TriLog sensor plug configuration.
When you define a custom sensor in the MultiLab program you must choose one
sensor from the list provided. Note its position in the list and look in Table 1 to find
the resistance.
Table 1: Auto recognition resistors
Position in
Define
sensor
dialog box
choose
list
Resistanc
e (k
Ω
)
Max
Resistance
(k
Ω
)
Min
Resistance
(k
Ω
)
1 372
384.1
360.9
2 410
424.8
396.5
3 455
472.9
437.8
4 508
530.9
486.6
5 573
601.9
544.8
6 653
691.1
615.6
7 754
806.5
703.4
8 888
961.9
814.6
9 1071
1183.0
959.2
10 1337
1524.4
1151.2
GND
Analog input
ID resistor
Sensor supply (+5V)