9 Measurement
FLUXUS F401
9.2 Measurement settings
61
UMFLUXUS_F401V2-1EN, 2020-06-12
Tab. 9.2:
Error output
error value
result
Minimum
the lower limit of the output range is output
Hold last value
the last measured value is output
Maximum
the upper limit of the output range is output
Other value...
The value has to be entered manually. It has to be within the limits of
the output.
Example
source item:
output:
output range:
volumetric flow rate
current output
4…20 mA
error delay: t
d
> 0
(see section 9.2.5 and Tab. 9.3)
The volumetric flow rate cannot be
measured during the time interval
t
0
...t
1
. The error value will be output.
Tab. 9.3:
Examples for the error output (output range: 4...20 mA)
list item
output signal
Minimum (4.0 mA)
v [m³/h]
t
0
t
1
t
I [mA]
20
4
t
d
t
0
t
1
t