7 Connection
7.3 Outputs
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
42
Removal
• Push the connector firmly toward the transmitter. Turn the knurled ring to the left and re
-
move the connector from the transmitter socket.
• Screw the protective cap on the socket. Turn protective cap to the right in order to fix it.
7.3
Outputs
Important!
→ Use only the supplied power supply unit.
→ The power supply is not protected against moisture. Use it only in dry rooms.
→ The voltage indicated on the power supply unit must not be exceeded.
→ Do not connect a defective power supply unit to the transmitter.
Tab. 7.1:
Circuit of the outputs
output
transmitter
external circuit
remark
internal
circuit
connection
passive
current
output
U
ext
=
4…24 V
U
ext
> 0.022
A
·
R
ext
[Ω] +
4
V
example:
U
ext
= 6 V
R
ext
≤ 90 Ω
R
ext
is the sum of all ohmic resistances in the circuit (e.g., resistance of the conductors, resistance of
the amperemeter/voltmeter).
C(+)
D(-)
+
-
mA
U
ext