9 Measurement
9.1 Parameter input
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
52
9.1.1
Input of pipe parameters
Outer pipe diameter/pipe circumference
• Enter the outer pipe diameter.
• Press ENTER.
An error message will be displayed if the entered parameter is outside the range. The
limit is displayed.
Example: upper limit 1100 mm for the connected transducers and for a pipe wall thick
-
ness of 50 mm.
It is possible to enter the pipe circumference instead of the outer pipe diameter, see
section 18.1).
If the input of the pipe circumference is activated and zero is entered in
Outer Diameter
,
the menu item
Pipe Circumfer.
will be displayed. If the pipe circumference is not to be en
-
tered, press key BRK to return to the main menu and start the parameter input again.
Pipe wall thickness
• Enter the pipe wall thickness.
• Press ENTER.
Parameter\Outer Diameter
Notice!
The inner pipe diameter (= outer pipe diameter - 2 × pipe wall thickness) will be cal
-
culated internally.
If the value is not within the inner pipe diameter range of the connected transducers,
an error message will be displayed.
Parameter\Wall Thickness
Outer Diameter
1100.0
MAXIMAL