9 Measurement
9.3 Start of the measurement
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
68
Input of the sound path number
A value for the number of sound paths corresponding to the connected transducers and
the entered parameters is recommended.
• Change the value, if necessary.
• Press ENTER.
Profile correction
If
With disturbance
is selected in the menu item
Special Funct.\SYSTEM settings\
Measuring\ProfileCorr 2.0
, it has to be checked whether the measurement arrangement
is appropriate.
If the number of sound paths is odd and more than one measuring channel is activated,
the following display appears:
• Select
no
if there are 2 transducer pairs in X arrangement or displaced X arrangement
at the measuring point (appropriate measurement arrangement). The profile
correction 2.0 at non ideal inflow conditions will be used. Transverse flow effects will be
compensated.
• Select
yes
if there is only one transducer pair at the measuring point (inappropriate
measurement arrangement). The profile correction 2.0 at non ideal inflow conditions
cannot be used. The profile correction 2.0 at ideal inflow conditions will be used.
Transverse flow effects will not be compensated.
• Press ENTER.
If
yes
is selected, the following menu messages are displayed:
Measuring\Channel\...\Sound Path
A: Alone at measp
>NO<
yes
Disturb correct.
not applicable!
I assume ideal
inlet conditions