9 Measurement
9.4 Display of measured values
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
74
9.4.2
Transducer distance
The transducer distance can be displayed during the measurement by scrolling the key
.
The optimum transducer distance (here: 51.2 mm) will be displayed in parentheses in the
upper line, followed by the entered transducer distance (here: 50.8 mm).
The optimum transducer distance might change during the measurement (e.g., due to
temperature fluctuations).
A deviation from the optimum transducer distance (here: 0.4 mm) will be compensated
internally.
F
flow velocity
comparison of the measured flow velocity with the flow limits of the system
√
OK, the flow velocity is not within the critical range
↑
the flow velocity is higher than the current limit
↓
the flow velocity is lower than the current cut-off flow
0
the flow velocity is within the limit range of the measuring method
?
unknown, cannot be measured
Fig. 9.4:
Display of the transducer distance
Notice!
Never change the transducer distance during the measurement.
Tab. 9.5:
Description of the status line
value
explanation
L=(51.2) 50.8 mm
54.5
m3/h