16 Advanced functions
16.11 Behavior of the alarm outputs
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
118
When adjusting to changed measurement conditions e.g, a substantial increase of the
fluid temperature, the alarm will not be switched. Alarm outputs with the switching condi
-
tion
OFF
will be set automatically to the switching function
NO Cont.
The alarm state can be displayed after the configuration of the alarm outputs and during
the measurement. This function is activated in
Special Funct.\SYSTEM settings\
Dialogs/Menus
. The activation of this function is recommended when alarm outputs
have to be reconfigured frequently.
• Select the menu item
SHOW RELAIS STAT
.
• Select
on
to activate the alarm state indication.
• Press ENTER.
If the alarm state indication is activated, the alarm output state is displayed after configur
-
ing the alarm outputs.
The alarm state indication is structured as follows:
Rx =
, with
X
being the number of the alarm output and
a pictogram
according to Tab. 16.3.
The configuration of the alarm outputs can be repeated by pressing key C. When the
configuration of the alarm outputs is finished, press ENTER. The main menu will be dis
-
played.
Fig. 16.2: Behavior of a relay when the flow direction changes
Notice!
There is neither a visual nor an acoustic indication of the alarm output switching.
Special Funct.\SYSTEM settings\Dialogs/Menus\SHOW RELAIS STAT
0
0
+
-
+
-
flow
flow
type
NON-HOLD
type
HOLD
approx. 1 s
reset of the alarm
(3 times key C)