17 SuperUser mode
FLUXUS F401
17.2 Defining flow parameters
121
UMFLUXUS_F401V2-1EN, 2020-06-12
17.2
Defining flow parameters
In the SuperUser mode, it is possible to define some flow parameters (profile bounds,
correction of the flow velocity) for the specific application or measuring point.
• Select
Special Funct.
\
SYSTEM settings
\
Measuring
\
Calibration
.
• Press ENTER.
17.2.1
Profile bounds
• Select
user
to define the profile bounds. If
factory
is selected, the default profile
bounds will be used and the menu item
Calibration
will be displayed.
• Press ENTER.
• Enter the max. Reynolds number at which the flow is laminar. The entered number will
be rounded to hundreds.
• Enter zero to use the default value of 1000.
• Press ENTER.
• Enter the min. Reynolds number at which the flow is turbulent. The entered number will
be rounded to hundreds.
• Enter zero to use the default value of 3000.
• Press ENTER.
A request is displayed if an additional correction of the flow velocity is to be defined.
• Select
on
to define the correction data,
off
to work without correction of the flow veloc
-
ity and return to the menu item
SYSTEM settings
.
Special Funct.\SYSTEM settings\Measuring\Calibration
Special Funct.\...\Calibration\...\Profile bounds
Special Funct.\...\Calibration\...\Laminar flow
Special Funct.\...\Calibration\...\Turbulent flow
Special Funct.\...\Calibration\...\Calibration