3 General principles
3.3 Acoustic penetration
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
22
Vertical pipe
Select the measuring point at a pipe location where the fluid flows upward. The pipe has
to be completely filled, see Fig. 3.11 and Fig. 3.12.
Free inlet or outlet pipe section
Select the measuring point at a pipe section where the pipe cannot run empty, see Fig.
Fig. 3.11: Recommended transducer
mounting position
Fig. 3.12: Disadvantageous transducer
mounting position
Fig. 3.13: Recommended transducer
mounting position
Fig. 3.14: Disadvantageous transducer
mounting position