6 Installation
6.2 Transducers
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
38
6.2.2.4
Mounting of the transducers with transducer shoes and ladder chains
• Adjust the transducer distance on the ruler before installation of the transducer shoes.
• Place the transducer shoes on the pipe.
• Loosen the knurled screw.
• Place the chain around the pipe.
• Tighten the chain and insert it into the hook of the transducer shoe.
• Tension the chain by tightening the knurled screw.
• Fix the second transducer shoe in the same way.
• Insert the transducer into the transducer shoe.
• Tighten the screw on the upper side of the transducer shoe in order to fix the transducer
on the pipe.
• Fix the second transducer in the same way.
• Read the transducer distance from the ruler.
Fig. 6.6:
Mounting of the transducers with transducer shoes and ladder chains
1 – ruler
2 – screw
3 – knurled nut
4 – transducer shoe
5 – hooks
6 – ladder chain
1
2
6
5
4
3