9 Measurement
9.2 Measurement settings
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
56
9.2
Measurement settings
9.2.1
Selection of the physical quantity and the unit of measurement
The following physical quantities can be measured:
• sound speed
• flow velocity: is calculated on the basis of the measured transit time difference
• volumetric flow rate: is calculated by multiplying the flow velocity by the cross-sectional
pipe area
• mass flow rate: is calculated by multiplying the volumetric flow rate by the operating
density of the fluid
The physical quantity is selected as follows:
• Select the program branch
Output Options
.
• Press ENTER.
• Select the physical quantity in the scroll list.
• Press ENTER.
For the selected physical quantity (except for the sound speed), a scroll list with the avail
-
able units of measurement is displayed. The unit of measurement which was selected
previously is displayed first.
• Select the unit of measurement of the selected physical quantity.
• Press ENTER.
Output Options
Physic. Quant.
Unit Of Measure
Damping
Output Options\Physic. Quant.
Output Options\Physic. Quant.\Volume flow
Notice!
If the physical quantity or the unit of measurement is changed, the settings of the
outputs have to be checked, see section 9.2.3.
Содержание FLUXUS F401
Страница 1: ...FLUXUS F401 UMFLUXUS_F401V2 1EN Operating instruction...
Страница 157: ...Annex FLUXUS F401 D Conformity declarations 157 UMFLUXUS_F401V2 1EN 2020 06 12 D Conformity declarations...
Страница 158: ...Annex D Conformity declarations FLUXUS F401 2020 06 12 UMFLUXUS_F401V2 1EN 158...
Страница 159: ......
Страница 160: ......