• Press ENTER.
9.2.5
Input of the error delay
The error delay is the time after which an error value will be sent to an output if no valid
measured values are available.
This display will only be indicated if the list item
edit
is selected in
Special
Funct.\Dialogs/Menus\Error-val. delay
.
If the error delay is not entered, the damping factor will be used.
• Enter a value for the error delay.
• Press ENTER.
Output Options\...\I1:Test output?
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.
: 0 m³/h
Full-Scale Val.
: 300 m³/h
Test value
= 150 m³/h (center of the measuring range, corresponds to 12 mA)
If the multimeter displays 12 mA, the current output functions correctly.
Output Options\...\I1:Error-val. delay