9 Measurement
FLUXUS F401
9.2 Measurement settings
65
UMFLUXUS_F401V2-1EN, 2020-06-12
• Enter the lowest expected measured value. The unit of measurement of the source item
will be displayed.
Zero-Scale Val.
is the value assigned to the value
Output MIN
of the output range.
• Press ENTER.
• Enter the highest expected measured value. The unit of measurement of the source
item will be displayed.
Full-Scale Val.
is the value assigned to the value
Output MAX
of the output range.
• Press ENTER.
Function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
yes
to test the output.
• Press ENTER.
• Enter a test value for the selected physical quantity. If the external measuring instru
-
ment displays the entered value, the output functions correctly.
• Press ENTER.
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.
: 0 m³/h
Full-Scale Val.
: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA
Output Options\...\I1:Test output?
Output Options\...\I1:Test value
Содержание FLUXUS F401
Страница 1: ...FLUXUS F401 UMFLUXUS_F401V2 1EN Operating instruction...
Страница 157: ...Annex FLUXUS F401 D Conformity declarations 157 UMFLUXUS_F401V2 1EN 2020 06 12 D Conformity declarations...
Страница 158: ...Annex D Conformity declarations FLUXUS F401 2020 06 12 UMFLUXUS_F401V2 1EN 158...
Страница 159: ......
Страница 160: ......