18 Settings
FLUXUS F401
18.2 Measurement settings
135
UMFLUXUS_F401V2-1EN, 2020-06-12
18.2
Measurement settings
• Select
Special Funct.\SYSTEM settings\Measuring
.
• Press ENTER.
• Select
yes
if the measured sound speed is to be compared to the theoretical or expect
-
ed value.
• Press ENTER.
The difference
δ
c = c
mea
- c
stored
between the two sound speeds will be displayed in the
lower line during the measurement. c
stored
is the sound speed of the reference fluid
stored in the database.
• Press key
during the measurement to scroll to the display of
δ
c.
• Select a list item:
–
off
: profile correction 1.0
–
on
: profile correction 2.0 at ideal inflow conditions (default)
–
With disturbance
: profile correction 2.0 at non ideal inflow conditions
• Press ENTER.
• Select
normal
to display and transmit the profile corrected flow values,
uncorr.
to
display and output the flow values without flow profile correction.
• Press ENTER.
For further information, see section 16.8.
Special Funct.\SYSTEM settings\Measuring
Notice!
The settings will be stored at the end of the dialog. If the menu item is quit by press
-
ing key BRK, the changes will not be stored.
Special Funct.\...\Measuring\Compare c-fluid
Special Funct.\...\Measuring\ProfileCorr 2.0
Special Funct.\...\Measuring\Flow Velocity