16 Advanced functions
16.8 Uncorrected flow velocity
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
110
16.8
Uncorrected flow velocity
For special applications, the uncorrected flow velocity might be of interest.
The profile correction for the flow velocity is activated in
Special Funct.\SYSTEM
settings\Measuring\Flow Velocity
.
• Select
normal
to display and output the flow velocity with profile correction.
• Select
uncorr.
to display and output the flow velocity without profile correction.
• Press ENTER.
If
uncorr.
is selected, each time the program branch
Measuring
is selected it will be
requested whether the profile correction is to be used or not.
If
no
is selected, the profile correction will be switched off. All physical quantities will be
calculated with the uncorrected flow velocity.
During the measurement, the designation of the physical quantity is displayed in capital
letters to indicate that the value is uncorrected.
• Press ENTER.
If
yes
is selected, the uncorrected flow velocity will only be used if the flow velocity is se
-
lected as physical quantity in the program branch
Output Options
.
All other physical quantities (volumetric flow rate, mass flow rate, etc.) will be determined
with the corrected flow velocity.
During the measurement, the designation of the physical quantity "flow velocity" is dis
-
played in capital letters to indicate that the value is uncorrected.
• Press ENTER.
In both cases, the corrected flow velocity can also be displayed.
Special Funct.\SYSTEM settings\Measuring\Flow Velocity
PROFILE CORR.
>NO<
yes
FLOW VELOCITY
2.60
m/s
PROFILE CORR.
>no<
YES