9 Measurement
FLUXUS F401
9.2 Measurement settings
57
UMFLUXUS_F401V2-1EN, 2020-06-12
9.2.2
Input of the damping factor
Each displayed measured value is a floating average of all measured values of the last x
seconds, with x being the damping factor. A damping factor of 1 s means that the mea
-
sured values are not averaged because the measuring rate is approx 1/s. The default
value of 10 s is appropriate for normal flow conditions. Values which fluctuate strongly
due to a higher flow dynamic, require a higher damping factor.
• Select the program branch
Output Options
.
• Press ENTER until the menu item
Damping
is displayed.
• Enter the damping factor.
• Press ENTER.
• Press key BRK to return to the main menu.
9.2.3
Installation of an output
If the transmitter is equipped with outputs, they have to be installed and activated before
they can be used:
• assignment of the physical quantity (source item) to be transmitted to the output by the
source channel, and the properties of the signal
• definition of the output behavior in case no valid measured values are available
• activation of the installed output in the program branch
Output Options
Output Options\...\Damping
Notice!
While the low-power mode is activated, the damping factor is max. 4 s.
SYSTEM settings
Proc. outputs
Install Output
Enable I1
Source item