17 SuperUser mode
FLUXUS F401
17.9 Manual reset of the totalizers
129
UMFLUXUS_F401V2-1EN, 2020-06-12
Total digits = Fixed to x digit
The number of decimal points is constant. The max value of the totalizer is reduced with
the number of decimal places.
17.9
Manual reset of the totalizers
• Select
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
3xC
clear totals
is displayed.
• Select
on
to activate the manual reset of the totalizers,
off
to deactivate it.
• Press ENTER.
decimal places
max. value
max. display
0
< 10
10
±9999999999
1
< 10
8
±99999999.9
2
< 10
7
±9999999.99
3
< 10
6
±999999.999
4
< 10
5
±99999.9999
Notice!
The number of decimal places and the max. value defined here only affect the dis
-
play of the totalizers.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\
3xC clear totals
Notice!
The manual reset of the totalizers is still active after the deactivation of the Supe
-
rUser mode.
Содержание FLUXUS F401
Страница 1: ...FLUXUS F401 UMFLUXUS_F401V2 1EN Operating instruction...
Страница 157: ...Annex FLUXUS F401 D Conformity declarations 157 UMFLUXUS_F401V2 1EN 2020 06 12 D Conformity declarations...
Страница 158: ...Annex D Conformity declarations FLUXUS F401 2020 06 12 UMFLUXUS_F401V2 1EN 158...
Страница 159: ......
Страница 160: ......