18 Settings
18.1 Dialogs and menus
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
132
18
Settings
18.1
Dialogs and menus
• Select
Special Funct.\SYSTEM settings\Dialogs/Menus
.
• Press ENTER.
18.1.1
Pipe circumference
• Select
on
if the pipe circumference is to be entered instead of the pipe diameter in the
program branch
Parameter
.
• Press ENTER.
If
on
is selected for
Pipe Circumfer.
, the outer pipe diameter will still be requested in
the program branch
Parameter
.
• In order to select the menu item
Pipe Circumfer.
, enter zero.
• Press ENTER.
The value displayed in
Pipe Circumfer.
is calculated on the basis of the last dis
-
played value of the outer pipe diameter.
Example: 100 mm
.
π = 314.2 mm
• Enter the pipe circumference. The limits for the pipe circumference are calculated on
the basis of the limits for the outer pipe diameter.
• Press ENTER.
During the next scroll through the program branch
Parameter
, the outer pipe diameter
that corresponds to the entered pipe circumference will be displayed.
Example: 180 mm : π = 57.3 mm
Special Funct.\SYSTEM settings\Dialogs/Menus
Notice!
The settings will be stored at the end of the dialog. If the menu item is quit by press
-
ing key BRK, the changes will not be stored.
Special Funct.\...\Dialogs/Menus\Pipe Circumfer.
Special Funct.\...\Dialogs/Menus\Pipe Circumfer.\Outer Diameter