Operating Procedures:
Optimizing Imaging
C O N F I D E N T I A L – FEI Limited Rights Data
5-13
Ion Beam tab
•
Quad
2D box – serves for minimization of image shift during
focusing.
•
L2 Modulator
button – starts automatic L2 voltage oscillation
(periodically under- and over-focuses imaging in a narrow
range) to facilitate the process.
•
Amplitude
slider – sets the modulation amplitude.
Ion Stigmator Balance tab
•
Stigmator Sin
/
Cos
2D box – serves for minimization of image
shift during astigmatism correction. The crosshair indicates
the actual stigmator setting.
Move the crosshair to suppress the imaging shift.
•
Modulator Sin
/
Cos
button – starts modulation of the
respective stigmator.
•
Amplitude
slider – sets the modulation amplitude.