Operating Procedures:
Specimen Preparation and Handling
C O N F I D E N T I A L – FEI Limited Rights Data
5-3
Specimen Baking Unit
The Specimen baking unit is intended to bake out the microscope specimens (up to 60 mm diameter) mounted on
the specimen holders.
C a u t i o n !
Read and follow all following instructions carefully!
Read and follow instructions stated in the FEI Microscope Systems Safety Manual carefully!
(installed with your Microscope control software)
Place instructions in a safe place for future reference.
Note the safety labels and engravings on the baking unit, read them and be careful to understand their meaning!
Do not allow anyone who has not read and understood these instructions and who cannot follow them to operate and /
or service the Specimen baking unit!
Failure to comply with the precautions and instructions provided with this unit, can result in bodily injury and property
damage from hazards of fire or burn.
W A R N I N G !
Do not leave the unit unattended when it is ON!
Do not open the lid when the unit is ON! Be extremely careful to avoid bodily injuries from burning.
When opening / closing the lid always use the plastic knobs, otherwise there is risk of burning!
When replacing the halogen bulb always wear protective gloves and safety googles!
W a r r a n t y d i s c l a i m e r
The unit must not be used for any other purposes than it is intended to otherwise the warranty claims cannot be
accepted.
FIGURE 5-1
Baking unit / Baking unit with opened lid
Installation instructions
The unit should be placed away from the flammable materials.
Do not put any objects on the top side of the unit – the unit becomes warm when in operation.
The unit is supplied with the voltage 12 V DC. The adaptor has a replaceable plug to fit in various mains sockets.
Connect the supply voltage cable connector to the unit and plug the adaptor to the mains voltage 110–230 V AC.
Содержание Scios 2
Страница 1: ...User Operation Manual Edition 1 Mar 2017 ...
Страница 84: ...Software Control Entering Commands in Summary C O N F I D E N T I A L FEI Limited Rights Data 3 58 ...
Страница 97: ...Alignments E Column Supervisor Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 13 Focus Centering ...
Страница 102: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 18 I Column Alignments ...
Страница 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Страница 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Страница 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Страница 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...