Operating Procedures:
Patterning
C O N F I D E N T I A L – FEI Limited Rights Data
5-49
•
ScanType
– the
Serpentine
means the beam proceeds from left
to right and back from right to left, while the
Raster
scans from
left to right, then the beam returns to the left starting point
•
Fill Style
– one can choose either to mill a
solid
(area) or just a
frame
(box and circular types only)
•
Passes –
a number of
the beam scans over the pattern
•
Defocus
of the beam (WD change) – influences the Total
Diameter and Area Calculation. It allows focusing above /
below (negative / positive value) the sample surface
•
Blur
– like Defocus, but specifying the (additional) diameter of
the blurred spot
•
Interaction Diameter
for an infinitely small beam – influences
the Total diameter
•
Total Diameter
– the combination of the beam diameter and
interaction diameter influences the OverlapX/Y and PitchX/Y
values (read only)
•
Maximum Dose per Area –
describes the adsorbed gas layer, allowing a certain dose to be deposited at a higher
rate than the saturation current density, allowing a temporary higher rate (actually not used)
•
Saturation Current Density
– the current at which 63% of the saturation sputter rate is reached (actually not used)
•
Total Volume Sputter Rate
– the speed at which material is removed or deposited (actually not used)
•
Selective Milling Enabled
/
Selective Milling Time Interval
/
Min Contrast Threshold
/
Max Contrast Threshold
items –
correspond to the
Selective Mill
tab module elements:
Enabled
check box /
Interval
adjuster / left / right border of
the grey level to be processed for the selected pattern
Selective Mill tab Module
It is possible to set a given grey level range to be processed in the
extent of the selected pattern area.
•
Scan
button – reads the pattern area grey level histogram
Move the vertical black line to a desired grey level position and
also move the red rectangle edges to set the grey level extent.
•
Enabled
check box – when ticked the selective milling is enabled
•
Interval
adjuster – sets the time [s] after which the histogram is actualised
Содержание Scios 2
Страница 1: ...User Operation Manual Edition 1 Mar 2017 ...
Страница 84: ...Software Control Entering Commands in Summary C O N F I D E N T I A L FEI Limited Rights Data 3 58 ...
Страница 97: ...Alignments E Column Supervisor Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 13 Focus Centering ...
Страница 102: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 18 I Column Alignments ...
Страница 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Страница 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Страница 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Страница 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...