Maintenance
:
Cleaning Procedures Overview
C O N F I D E N T I A L – FEI Limited Rights Data
6-2
W A R N I N G !
The cleaning solvents ethanol and isopropanol are highly flammable! Do not use open flames and do not smoke while
cleaning. Ventilate the room properly.
Cleaning Column Parts
All column parts are polished before the instrument is delivered. For this reason only occasional light polishing is
required to remove contamination that may build up on components in the column and specimen chamber as part
of normal operation. Any part that is exposed to the electron beam should be highly polished, and free of
contamination and / or scratches that can charge and thus degrade the image.
C a u t i o n !
Final lens pole is allowed to clean only from the outside! There is a risk of T1 detector damage.
Gold plated parts should not be polished with abrasive.
Materials and Technique
To polish components, place a lint-free cloth on a flat surface (a glass block is ideal) and apply a small amount of
Soft Scrub or CIF and distilled water to the cloth.
Place the part to be cleaned on the polish and rub with a circular motion until all contamination is removed. For
inner surfaces, use a cotton swab or wooden dowel as an applicator. A toothpick can be used for small holes.
Lint-free nylon (not cotton) or latex surgical gloves should be worn while handling parts to avoid contaminating
just-cleaned surfaces. Tweezers should be used to hold small parts.
After the part has been polished, remove the Soft Scrub/CIF cleaner by washing in hot water. Inspect the part
under a stereo microscope at 20× magnification to ensure that there is no remaining contamination or polish
residue. Wash the part in de-ionized or distilled water in a beaker with an ultrasonic cleaner for several minutes.
Transfer the part to a clean beaker with alcohol or isopropanol and clean ultrasonically again for several minutes.
When the components are dry (a compressed air “duster” can speed drying), reassemble and return them to the
column. If a part is stained, heat it with hot water and immediately rinse with alcohol and dry using compressed air.
Cleaning Tips
Parts exposed to the electron beam require periodic polishing. This will ensure maximum performance of the
instrument for many years.
Do not use metal polishes such as POL or WENOL to clean parts as these can leave outgassing material. Be aware
that threaded surfaces should not be polished as these do not have contact to the beam and are a source of
outgassing if polish is trapped. Wash threads with alcohol or isopropanol if absolutely necessary.
After cleaning, inspect all parts for residue and stains using a light microscope.
Table 6-1
Household Cleaners
Country
Name
Country
Name
Austria
CIF
Italy
CIF
Australia
CIF
Japan
CIF
Finland
CIF
Netherlands
CIF
France
CIF
Switzerland
CIF
Germany
CIF
UK
CIF
USA
Soft Scrub
Содержание Scios 2
Страница 1: ...User Operation Manual Edition 1 Mar 2017 ...
Страница 84: ...Software Control Entering Commands in Summary C O N F I D E N T I A L FEI Limited Rights Data 3 58 ...
Страница 97: ...Alignments E Column Supervisor Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 13 Focus Centering ...
Страница 102: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 18 I Column Alignments ...
Страница 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Страница 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Страница 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Страница 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...