Software Control
:
Microscope Control Software
C O N F I D E N T I A L – FEI Limited Rights Data
3-16
Unlink Z to FWD
This feature functions in the opposite way as the following one.
The Z coordinate value represents then the distance from the Z
home position (stage base). The dialog warns you about the
stage Z-axis positive move direction.
Link Z to FWD (Ctrl + L)
This functionality provides the system with definite stage
positioning to perform accurate and secure movement between
the sample top surface and the end of the objective lens. It sets
the Z coordinate value to the actual Free Working Distance
(FWD) value. The related toolbar icon changes according to the
Z-coordinate status:
•
Greyed icon
: the function is disabled – the tooltip gives a
reasons (for instance high voltage is switched off, all displays
are paused, beam is not on…).
•
Red question mark
: the function is enabled – Z is not linked to FWD. Use this function as soon as possible, after a
proper focus.
•
Red circle arrow
: the function is enabled – Z is roughly linked to FWD, but it needs correction. It happens e.g. after:
pumping the chamber when specimens were exchanged, focusing and linking Z to FWD at a long WD and then
moving the stage to a short WD. In this case focus carefully at a WD around 7 mm and use this function again.
•
Green 2-ended arrow
: the function is enabled – Z is properly linked to FWD. Now it should be safe to change the
working distance by setting the Z coordinate in the Stage module.
Note
The Do not show this message again check box disables this notice once for ever.
Enable Z-Tilt Safety Map
Some movements of the tilted stage are not safe because of a possible collision with the final lens. The table (not
user editable) with pairs of values indicates the maximum safe Tilt angle for a certain Z value when linked. It can be
used to guarantee safe usage of the stage (tilt restriction) for flat samples only and when a proper Link Z to FWD is
established.
Enable Safe Stage Moves
When ticked the software takes into account all accessories installed inside the chamber. That’s why the stage
doesn’t move to the target position following the straight path. When the magnification is large and stage
movements are expected to be tiny, it is not necessary to switch on this functionality.
Tilt 0° (Ctrl + E) / Tilt 52° (Ctrl + I)
tilts the stage to 0° / 52°.
Sample Navigation (Ctrl + N)
toggles on / off function that enables to navigate electron / ion imaging (scan field) towards desired places on a
specimen using either paused or loaded sample image (usually captured at much lower magnification).
The Sample Navigation can be selected independently for any display, regardless of its
actual content and status. A tick next to the menu item indicates that the function is
selected for the selected display. As soon as this imaging is paused, the Sample Navigation
indicator appears in the upper right corner of the display. The indicator is green as long as
the paused imaging can be used to navigate the live one, otherwise turns red.
Navigation Montage
This procedure enables to capture the sample Navigation image to be used in the Sample Navigation.
Navigation Alignment
This procedure aligns the Navigation image according to the live reference image.
Содержание Scios 2
Страница 1: ...User Operation Manual Edition 1 Mar 2017 ...
Страница 84: ...Software Control Entering Commands in Summary C O N F I D E N T I A L FEI Limited Rights Data 3 58 ...
Страница 97: ...Alignments E Column Supervisor Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 13 Focus Centering ...
Страница 102: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 18 I Column Alignments ...
Страница 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Страница 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Страница 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Страница 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...