Can the SpiroSphere system run without network access?
Yes. Alternatively, the data can be uploaded using a USB stick on a local Internet PC.
Speci
¿
cation of network connection
The SpiroSphere can be connected to an IT network in order to send a pdf report to a designated
email address. In order to use this functionality, the SpiroSphere must be connected to an IT-
network with TCP/IP connectivity via Ethernet, WLAN (or optional 3G) with routing to the Internet.
The con guration of the necessary network settings is described in the "Settings and Tools" menu
in the sub menu "Communication".
Add description and the con guration of the electronic report functionality can be found in the
section "Report & Printing".
Safety hints
The integration of the SpiroSphere to an IT-network that includes other devices can incorporate
risks to patient, operator or other persons. The user/ operator who integrates the SpiroSphere to
the network is responsible to analyze, rate and handle those risk accordingly (eg. according IEC
80001-1).
The following changes in the IT-network might lead to new risks and therefore it might be
necessary that the user needs to reevaluate the analysis:
•
Change of the infrastructure
•
Integration of additional devices to the network
•
Removal of device from the network
•
Update of devices connected to the IT-network
•
Upgrade of devices connected to the IT-network
If you have technical questions or require further assistance please feel free to contact our
customer care.
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...