SpiroSphere
®
ECG Instructions for Use
Page 62/136
Version 02.02 • 04AUG2022
ECG Leads
To minimize artifacts, the skin of the de ned lead positions has to be prepared thoroughly.
Preparing the sub
M
ect
µ
s skin:
1. Identify the (10) electrode sites on the torso by referring to the picture and description below.
2. Remove any hair from the electrode site using a razor.
3. Wipe oils from the electrode sites with an alcohol prep pad.
4. Remove any dead skin from the electrode sites with an abrasive cleaner.
Two to three moderate rubs at each site should be suf cient.
TIP: Electrodes should be stored in an air-tight container. Electrodes will dry out
if not stored properly which will cause loss of adhesion and conductivity. Please
note the storage conditions indicated on the electrode packaging.
Correct electrode placement is important for acquiring a successful ECG recording.
Chest ECG
V1
Fourth intercostal space at the right
sternal
border
V2
Fourth intercostal space at the left
sternal
border
V3
Midway between V2 and V4
V4
Fifth intercostal space at the left of the
midclavicular
line
V5
Anterior axillary line at same horizontal
level as V4
V6
Mid-axillary line on same horizontal
level as V4 and V5
LA Left
wrist
RA Right
wrist
RL Right
ankle
LL Left
ankle
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...