Screen display after the rst effort:
Start the next effort by tapping on <
Start Effort
>.
If necessary, it is possible to terminate the test prematurely.
In this case, a warning message is displayed.
* Literature:
MR Miller et al. Series “ATS/ERS Task Force: Standardisation of Lung Function Testing“,
Standardisation of Spirometry, Eur Respir J 2005; 319-338. Copyright © ERS Journals Ltd. 2005
Scroll down to display further parameters
(if applicable)
The upper left part of the chart
section shows the recording of the
ow-volume loop. The upper right
part shows the volume-time tracing.
The lower section of the screen
displays the predicted values
calculated from the patient data and
the actual values measured from
rst effort.
Pred
=
Predicted
value
Best
= Best value of all
valid
efforts.
%Pred
= Best value in % of
predicted
values
Screen display after three efforts:
The "
Best
" column displays the
best value out of all valid efforts.
De nition of the best effort depends
on the Settings selected
(see: >Settings Spirometry >Forced
Spirometry >Measurement).
The
T
uality of the
À
ow
volume loop depends on the patient
µ
s cooperation.
In
order to assess repeatability and quality, it is recommended to perform at least 3 efforts
according to ATS/ERS guidelines.
The results of the best and the second best effort for FEV1 and FVC may differ by
150 mL. For FVC 1 L a difference of 100 mL is valid*.
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...