SpiroSphere
®
ECG Instructions for Use
Page 11/136
Version 02.02 • 04AUG2022
4. Prior to the rst usage, switch on the SpiroSphere Sensor
by pressing the "
Power On
" switch located at the back side
of the SpiroSphere Sensor for at least 3 seconds.
ERT PT with mouthpiece
ERT PT is only for single use. Do not reuse the ERT PT due to risk of cross
contamination.
Do not remove the mouthpiece from the ERT PT. Only use the ERT PT with
connected mouthpiece.
Prior to the rst use, make sure that the SpiroSphere Sensor is fully
charged.
5. Ensure an ERT PT is inserted correctly into the
SpiroSphere Sensor
(refer to Chapter "Sensor Insert").
Take care that you have aligned the rail with the sensor
holes of the PT tube with the grooved edge of the
SpiroSphere Sensor (as below) when inserting the ERT PT
into the SpiroSphere Sensor.
The maximum surface temperature of the SpiroSphere Sensor Unit can get
up to 47°C during charging.
Spirometry should only be performed by patients who can cooperate in the
performance.
Shut down
1. To power off the
SpiroSphere Main Unit
, press the "
Power
" button located at the front
of the main unit. Disconnect the main unit from the power source (please note that this
will prevent the battery from charging).
2. Power off the SpiroSphere Sensor by pressing the "
Power
" button located near
the LEDs of the SpiroSphere Sensor.
Do not position the Power Supply and the SpiroSphere so that it is dif cult
to operate the disconnection of the device from the mains supply.
rail with sensor holes
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...