The signals emitted by the measuring device do not comply with the alarmstandard
for medical electrical devices, DIN EN 60601-1-8.
All parts of the medical device, including the equipment, which come into contact
with the patient when operating the device in accordance to the regulations, meet the
applicable biocompatibility standards.
Packaging material can pose a risk of suffocation. Keep it inaccessible for children or
dispose of it in accordance with the regulations.
The medical device, especially the electrodes and cables, must only be used on
healthy and intact skin.
Do not make any changes to the electrodes or the device during operation. That
might result in incorrect measurements.
The device is not suitable for heart leads. When using a de brillator the electrodes of
ECG and de brillator must not come into contact.
If a de brillator is used, pay attention that no one has contact with the patient. Burns
or other injuries might result.
It is not possible to operate the medical device in combination with HF devices, e. g.
surgical units.
When monitoring critical patients, an alternative medical device must be kept
available at all times in case of a device breakdown.
Avoid tensional load on the cables.
Only use biocompatible and FDA or CE-certi ed electrodes, as applicable, with the
medical device.
Comply with the instructions for use of the applied electrodes at all times.
Assure that the electrodes are correctly applied. Incorrectly applied electrodes can
adulterate the ECG and result in misinterpretations.
The skin at the application positions of the electrodes has to be clean and dry.
Moisture can result in signal distortion. If using an in ammable skin cleansing agents,
the skin must be dried completely as a means of precaution.
Conductive parts of the electrodes have to be kept away from other combustible
parts.
Electric Shock from Unconnected Electrodes:
As soon as at least one electrode
or peripheral sensor is attached to a person, make sure that none of the other
electrodes or peripheral sensors come into contact with any conductive material that
might provide a return path to ground.
Operation Cautions and Warnings
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...