SpiroSphere
®
ECG Instructions for Use
Page 122/136
Version 02.02 • 04AUG2022
Electromagnetic Emission and
Immunity SpiroSphere ECG
Medical electrical equipment is subject to special precautions regarding electromagnetic
compatibility (EMC). User must follow the EMC installation instructions and only use the device in
the intended environment during operation.
The SpiroSphere meets the regulations according to EN60601-1-2 regarding the interference
radiated and received. For more information, refer to the EMC tables of the instructions for use of
your device/system.
The SpiroSphere is intended to be used in professional healthcare facility environment like
doctor’s of ce or hospital.
SpiroSphere conforms to basic standard IEC 60601-1.
The essential performance of the device is the acquisition of spirometry and ECG signals and
display on the Main Unit.
Electromagnetic irradiation can influence the device essential performance by the
following:
• The maximum error (drift, uncertainty) of Flow measurements might be up to +/- 20ml/s
Electromagnetic irradiation can influence the ECG device essential performance by the
following:
• The measurement signal might be in uenced but will recover for use within 10s after the source
of disturbance disappeared.