SpiroSphere
®
ECG Instructions for Use
Page 110/136
Version 02.02 • 04AUG2022
The connection of further power-operated units to your
ERT
unit may cause all
the leakage currents to add-up and the safety of the subject is reduced. Due to
this, the connection of further units may only be carried out on consultation with
the
ERT
Customer Care.
Accessory equipment connected to the interfaces must be certi ed according to
the respective EN standards (e.g. EN 60950 or EN 60601-1).
Everybody who connects additional equipment to the signal input part or signal
output part con gures a medical system, and is therefore responsible that
the system complies with the requirements of the valid version of the system
standard (IEC 60601-1). If in doubt, consult the technical service department or
your local representative.
Use of Multi
Plug Sockets
Do not connect the medical device/system to multi-plug sockets.
Radiated Interference
The ERT device meets the regulations according to EN 60601-1-2 (CISPR
11 Group 1 class B) regarding the interference radiated and received. The
device should not be installed in the vicinity of high-frequency devices, X-ray
equipment, motors or transformers with high installed power rating since electric
or magnetic interference elds may falsify the result of measurements or make
taking measurements impossible. Due to this, the vicinity of power lines is to be
avoided as well. Existing environmental interferences may cause deviations of
the measurement values without impairing the device’s function. Therefore, it is
recommended to keep a distance of about 2 meters from possible error sources
when using the device.
This device should not be operated in immediate vicinity to or stacked with other
devices since this could lead to an incorrect operation. However, if an operation
in the described manner is necessary, this device as well as the other devices
should be carefully observed to ascertain a proper operation.
Using other accessories, other transformers and other cables than those
speci ed or provided by the device's manufacturer can result in increased
electromagnetic radiation or reduced electromagnetic immunity of the device
and can lead to an incorrect operation.
Portable RF communications equipment (transmitters) (including appropriate
accessories such as aerial wires and external antennas) should be operated
with a minimum distance of 30 cm (21 inch) to the SpiroSphere's components
and cables speci ed by the manufacturer. Non-observance may lead to a
reduction of the device's performance.
For further information see chapter
Electromagnetic Emission and Immunity
and
Notes on EMC.
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...