SpiroSphere
®
ECG Instructions for Use
Page 112/136
Version 02.02 • 04AUG2022
Cleaning and Hygiene
Prior to every application, all parts which come in contact with the patient and
which are intended for reuse must be cleaned or disinfected (unless otherwise
instructed).
Prior to taking measurements of a patient, his/her medical history is to be
checked in order to avoid a contamination of the device and a resulting cross-
contamination of the next patient.
While performing a calibration check, a new disposable Pneumotach must be
used to prevent cross-contamination between the calibration syringe and the
parts. This will prevent contamination of the syringe and allow for its reuse.
Always be sure to disconnect the devices/systems from the mains power before
cleaning or disinfecting. The Main Unit corresponds to protection class IP21, the
SpiroSphere Sensor to class IP20.
The device may not be soaked in liquid of any kind. Liquid inside the device/
system may lead to harm of the user and can destroy the device.
The device can be cleaned with a damp (but not soaked) cloth, which does not
produce lint. More detailed information can be found under “Hygiene“ in this
Instructions for Use.
Detergents and chemicals required for cleaning and disinfection must always be
stored in specially marked containers to prevent any accidental improper use.
Biocompatibility
Component
Material
Mouth piece
Styrolution PS 454N HIPS
Housing parts of the sensor unit and
the main unit
Cycoloy CX2244ME
PT-tube
Styrolution PS 454N HIPS
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...