SpiroSphere
®
ECG Instructions for Use
Page 13/136
Version 02.02 • 04AUG2022
Problem
Troubleshooting
SpiroSphere Main Unit
does not start
SpiroSphere Main Unit is not switched on. Try to switch on with the
power button (press the button for at least 3 seconds).
Power supply is not connected and the battery discharged. Connect
the power supply and let the SpiroSphere Main Unit charge for a
few minutes.
SpiroSphere Main Unit is in standby mode. Use the touch screen or
power button to wake up.
SpiroSphere Sensor does
not start
SpiroSphere Sensor is not switched on. Press the power button for
at least 3 seconds to switch the SpiroSphere Sensor on.
Battery is discharged. Place the SpiroSphere Sensor in the
charging cradle for at least 10 minutes, then try again.
SpiroSphere Main Unit
LED is blinking orange
Battery is discharged. Connect power supply immediately.
SpiroSphere Sensor
orange LED is blinking
slowly
Battery is discharged. Place SpiroSphere Sensor in charging cradle
of the SpiroSphere Main Unit.
SpiroSphere Sensor LED
is off while in the charging
cradle
SpiroSphere Sensor is in sleep mode. Remove from the cradle and
insert again.
SpiroSphere Sensor might be fully charged or off.Try to identify
the SpiroSphere Sensor in the Spirometry Settings. Place the
SpiroSphere Sensor again in the charging cradle.
SpiroSphere Sensor can't
be paired
SpiroSphere Sensor is powered off. Switch on SpiroSphere Sensor
(see above)
SpiroSphere Sensor is not in the pairing mode. Remove the
SpiroSphere Sensor from the charging cradle if already inserted
and then insert the SpiroSphere Sensor in the charging cradle.
SpiroSphere Sensor is now in pairing mode for 90 seconds.
Connection to the
SpiroSphere Sensor is
not possible
SpiroSphere Sensor is off. Power on SpiroSphere Sensor (see
above).
SpiroSphere Sensor is out of range. Check that the SpiroSphere
Sensor is in the range (10m in the line of sight).
Calibration Check or
Linearity Check not
successful
Check if the calibration pump has the correct volume setting.
Check if ERT PT is inserted correctly.
ERT PT has been used – replace by a new ERT PT
SpiroSphere Main Unit is
not responding
Touch the Home icon (top left).
Restart the SpiroSphere Main Unit by pressing the Power button for
3 seconds – Shutdown Menu – use restart.
Press the power button for more than 8 seconds to switch off the
SpiroSphere Main Unit.
Then restart by pressing the power button for min 3 seconds.
Troubleshooting
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...