Version 02.02 • 04AUG2022
Recurrent Test
Medical Electrical equipment needs a recurrent testing after repair of the
equipment according to IEC 62353.
The calibration syringe itself has to be calibrated at regular intervals as
determined by the manufacturer and as indicated on the syringe. The calibration
syringe must be checked for an accuracy of ± 12 mL.
Recycling
Adhere to the national law of the country when disposing the medical device
and its accessories.
Improper disposal of the device and/or its accessories can result in serious
environmental hazards.
Maintenance
No part of the medical device may be replaced by the customer.
Use only
ERT
approved accessories and spare parts for this medical device.
If applied parts (e.g. SpiroSphere Sensor) have been exposed to extreme
mechanical stress, a function test (e.g. volume calibration check) has to be
performed. If function is lost, the defective part is to be replaced. Damaged
parts, e.g. frayed plugs, receptacles, a damaged handle, and defective cables
should be replaced immediately by an authorized specialist or engineers from
ERT Customer Care.
The device must not be opened. If it is opened without
authorization the guarantee entitlement expires.
ERT Customer Care
is always
at your disposal with help and assistance in case of problems.
Before turning on the device/system you should always check whether the
power cable, power plug, outlet and power input of the device are free from
defects.
Before turning on the device/ system the following issues have to be checked
visually on a daily basis:
-
the display glass is undamaged
-
the unit has not been mechanically stressed in the extreme (e.g. damage
to the housing, the cable is made defective by running over it with a heavy
object or dragging it)
-
no liquid got inside the unit
-
the SpiroSphere Sensor is not damaged
-
cables and/or multiple connectors are not defective
-
coverings are not broken
An unattended child should
not
get into contact with disposables, accessories
and packing material as well as cleaning and disinfection substances.
Further safety-related checks in accordance with IEC 62353 have to be
performed every 3 years by an authorized technician. These safety-related
checks include the measuring of leakage current and insulation resistance and
the visual checks as performed on a daily basis.
If you recognize or experience any serious incident that has occurred in relation to the
device, you should report this to the manufacturer and the competent authority of your
country.
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...