SpiroSphere
®
ECG Instructions for Use
Page 131/136
Version 02.02 • 04AUG2022
Feature
Value
Classi cation according to 60601-1
- Protection type against electric shock
- Protection level against electric shock
Device with internal power supply
Type CF
Electromagnetic compatibility (EMC)
according to 60601-1-2
- Noise suppression
- Immunity
EN 55011
EN 61000-4 parts 2, 3, 6, 8
Protection level against access to hazardous
parts and ingress of solid objects
IP20
Variants
12-channel EU cable; 12-channel US cable,
12-channel connector
Accuracy of the ECG signal
1,94 V/bit resolution (at sampling rate of 500Hz)
Filter
bandpass 0,05 Hz – 150 Hz
no line lter (50Hz / 60Hz)
Electrodes
Standard ECG electrodes
Connection of the electrodes
4 mm snap, gilded
Shelf life
1 year for device
Product life
The product life is 5 years.
Radio transmission
Approved in accordance to RED directive
transmitter module marked by CE, manufactured
by Panasonic incorporated to OEM product.
Application
ECG Recording
ECG leads
acc. to Einthoven, Wilson, Goldberger
Bandwidth
0 - 150 Hz digital
Sampling rate per channel
500 Hz
Pacemaker detection
4000 Hz sampling rate
HR and pacemaker detection in ECG unit
Resolution
1.94
μ
V/bit ECG, 15 bit
Power supply
1 x AA battery
Connection to electrodes
4 mm snap connector, gold-plated
Moisture protection
IP20
Electrode cable length
0.44 m / 0.81 m
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...