SpiroSphere
®
ECG Instructions for Use
Page 103/136
Version 02.02 • 04AUG2022
Functional Check ECG
At each power-on, the medical device carries out an integrated self test. By means of this test the
correct functionality of the following components is ensured:
•
Voltage
supply
•
Memory
•
Bluetooth
module
•
ADC for pacemaker detection
•
ADC for ECG
An error condition in the Bluetooth connection is indicated by a quick double ashing of the blue
LED. Error conditions of all other components are transmitted via the Bluetooth interface to the
host PC.
Conduct a visual inspection of the medical device before each use. In case
you should discover external damage at the device or the cables or in case
the integrated self test fails, do not continue to operate the device. Contact
the manufacturer Corscience or a specialist distributor authorized by
Corscience to have the device repaired.
Troubleshooting Guide ECG
The distance between the ECG unit and the SpiroSphere main unit
impacts the quality of the Bluetooth connection. Keep it as short as
possible and free of objects between both, e.g. furniture. If Bluetooth
transmission fails for more than 10 seconds, the old ECG data are
dismissed by the device and therefore lost for the recording.
The separation of the point-to-point Bluetooth connection from another
network (e.g. intranet, internet) is under the responsibility of the
operator. In case the point-to-point Bluetooth connection is not served
for more than 10 minutes the connection is terminated. The not yet
transferred ECG data from ECG unit will be lost after 10 seconds (see
above).
LED Status Chart
LED code (blue)
System status
Possibility of intervention
Blue LED blinks slowly (1Hz)
Bluetooth connection active
-
Blue LED slow double blink
Battery almost empty
Replace battery
Blue LED blinks fast (3Hz)
Bluetooth pairing active
ECG unit can be paired from the
according Settings menu
Blue LED on (solid)
Ready for operation
Start measurement in order to
establish connection
Blue LED quick double blink
Start-up error,
connection impossible
Replace battery, retry connection.
If continuous failure contact ERT
Customer Care
Содержание SpiroSphere
Страница 1: ...781239 Version 02 02 Instructions for Use SpiroSphere ECG ...
Страница 39: ...SpiroSphere ECG Instructions for Use Page 39 136 Version 02 02 04AUG2022 Pre Post Report ...
Страница 134: ...SpiroSphere ECG Instructions for Use Page 134 136 Version 02 02 04AUG2022 For your notes ...
Страница 135: ...SpiroSphere ECG Instructions for Use Page 135 136 Version 02 02 04AUG2022 ...