Proline t-mass 65 PROFIBUS DP
Trouble-shooting
81
No. # 6xx
→
Simulation mode active
601
S: POS. ZERO-RET.
! # 601
0x53
UNCERTAIN
Sensor
conversion
not accurate
Constant
Positive zero return
active
Cause:
Positive zero return is active.
!
Note!
This message has the highest display priority!
Remedy:
Switch off positive zero return.
Access:
SYSTEM PARAMETER
→
POS. ZERO RETURN
(
→
OFF)
691
S: SIM. FAILSAFE
!: # 691
0x48
0x49
0x4A
UNCERTAIN
Substitute set
O.K.
Low
High
Simulation failsafe
active
Cause:
Simulation of failsafe mode (outputs) is active.
Remedy:
Switch off simulation.
Access:
SIMULAT. SYSTEM
→
FAILSAFE MODE
(
→
OFF)
692
S: SIM. MEASURAND
!: # 692
0x60
0x61
0x62
UNCERTAIN
Simulated
Value
O.K.
Low
High
Simulation
measured value
active
Cause:
Simulation of the measured value is active.
Remedy:
Switch off simulation.
Access:
SIMULAT. SYSTEM
→
SIM. MEASURAND
(
→
OFF)
698
S: DEV. TEST ACT.
!: # 698
0x60
0x61
0x62
UNCERTAIN
Simulated
Value
O.K.
Low
High
Device test via
Fieldcheck active
Cause:
The measuring device is being checked on site via
the test and simulation device.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Remedy / spare parts
Qu
a
lit
y
c
o
d
e
(H
EX
)
Me
as
ure
d
va
ri
abl
e
s
tatus
Q
u
al
it
y
s
tatus
Q
u
al
it
y subs
tatus
Li
m
its