Proline t-mass 65 PROFIBUS DP
Trouble-shooting
79
215
S: LOAD S-DAT
$
: # 215
0x0F
BAD
Device Failure
Constant
Save to S-DAT
failed
Cause:
Data backup (downloading) to HistoROM/S-DAT
failed, or error when accessing (uploading) the
calibration values stored in the HistoROM/S-DAT.
Remedy:
1.
Check whether the S-DAT is correctly plugged
into the amplifier board
2.
Replace the S-DAT if it is defective. Spare parts
Page 84 Before replacing the DAT, check
that the new, replacement DAT is compatible
with the measuring electronics. Check the:
- Spare parts set number
- Hardware revision code
3.
Replace measuring electronics boards if
necessary. Spare parts
216
S: SAVE S-DAT
$
: # 216
0x0F
BAD
Device Failure
Constant
Restore from
S-DAT failed
251
S: COMMUNIC. SENS
$
: # 251
0x0F
BAD
Device Failure
Constant
Communication
sensor failure
Cause:
Internal microprocessor communication fault on
the amplifier board.
Remedy:
Remove the amplifier board. Spare parts
261
S: COMMUNICAT. I/O
$
: # 261
0x18
0x19
0x1A
BAD
No
Communica-
tion
O.K.
Low
High
Communication
I/O failure
Cause:
No data reception between amplifier and I/O board
or faulty internal data transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders
262
S: COMMUNICAT. I/O
$
: # 262
0x18
0x19
0x1A
BAD
No
Communica-
tion
O.K.
Low
High
Communication
I/O failure
Cause:
No data reception between amplifier and I/O board
or faulty internal data transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders
No. # 3xx
→
System limits exceeded
372
S: DIFF. TEMP. LOW
$
: # 372
0x13
BAD
Sensor Failure
Constant
Diff. Temp. is
below limit
Cause:
The measured sensor differential temperature is
below limit value.
Remedy:
Reduce the flow rate or consider replacing the
instrument with a suitable size for the application if
possible.
379
S: GAS MIX. CORR.
!: # 379
0x40
0x41
0x42
UNCERTAIN
O.K.
O.K.
Low
High
Gas mixture data
corrupted
Cause:
The defined gas mixture has been corrupted.
Remedy:
Gas mixture needs to be re-entered.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Remedy / spare parts
Qu
a
lit
y
c
o
d
e
(H
EX
)
Me
as
ure
d
va
ri
abl
e
s
tatus
Q
u
al
it
y
s
tatus
Q
u
al
it
y subs
tatus
Li
m
its
Содержание Proline t-mass 65 PROFIBUS DP
Страница 2: ......
Страница 46: ...Operation Proline t mass 65 PROFIBUS DP 46 ...
Страница 106: ...www endress com worldwide BA113D 06 en 02 06 71021608 FM SGML6 0 ProMoDo ...