Proline t-mass 65 PROFIBUS DP
Trouble-shooting
77
041
S: TRANSM. HW-DAT
$
: # 041
0x0F
BAD
Device Failure
Constant
T-DAT failure
Cause:
1.
HistoROM/T-DAT is defective.
2.
HistoROM/T-DAT is not plugged into the
amplifier board or is missing.
Remedy:
1.
Replace the HistoROM/T-DAT.
Spare parts
Check the spare parts set number to ensure
that the new, replacement DAT is compatible
with the measuring electronics.
2.
Plug the HistoROM/T-DAT into the amplifier
board
042
S: TRANSM. SW-DAT
$
: # 042
0x0F
BAD
Device Failure
Constant
T-DAT data
inconsistent
Cause:
Error accessing the calibration values stored in the
HistoROM/T-DAT.
Remedy:
1.
Check whether the HistoROM/T-DAT is
correctly plugged into the amplifier board.
2.
Replace the HistoROM/T-DAT if it is
defective.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
– Spare parts set number
– Hardware revision code
3.
Replace measuring electronics boards if
necessary.
051
AC COMPATIB.
$
: # 051
0x0F
BAD
Device Failure
Constant
Compatibility Amp.
I/O module
Cause:
The I/O board and the amplifier board are not
compatible.
Remedy:
Use only compatible modules and boards.
Check the compatibility of the modules used.
Check the:
– Spare parts set number
– Hardware revision code
052
S: HW-COMMODUL
$
: # 052
0x0F
BAD
Device Failure
Constant
I/O module
hardware failure
Cause:
The I/O board is defective.
Remedy:
Replace the I/O board. Spare parts
070
S: SENSOR DEFECT
$
: # 070
0x13
BAD
Sensor Failure
Constant
Sensor failure
Cause:
Flow sensors are likely to be defect, measurement
is no longer possible.
Remedy:
Contact your Hauser representative.
071
S: SENSOR DRIFT
$
: # 071
0x13
BAD
Sensor Failure
Constant
Sensor drift
Cause:
Calibration drift has been detected.
Remedy:
Contact your Hauser representative.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Remedy / spare parts
Qu
a
lit
y
c
o
d
e
(H
EX
)
Me
as
ure
d
va
ri
abl
e
s
tatus
Q
u
al
it
y
s
tatus
Q
u
al
it
y subs
tatus
Li
m
its
Содержание Proline t-mass 65 PROFIBUS DP
Страница 2: ......
Страница 46: ...Operation Proline t mass 65 PROFIBUS DP 46 ...
Страница 106: ...www endress com worldwide BA113D 06 en 02 06 71021608 FM SGML6 0 ProMoDo ...