Proline t-mass 65 PROFIBUS DP
Trouble-shooting
87
Wall-mount housing
#
Warning!
• Risk of electric shock. Exposed components carry dangerous voltages. Make sure that the power
supply is switched off before you remove the cover of the electronics compartment.
• Risk of damaging electronic components (ESD protection). Static electricity can damage electronic
components or impair their operability. Use a workplace with a grounded working surface
purposely built for electrostatically sensitive devices!
• If you cannot guarantee that the dielectric strength of the device can be maintained during the
following steps, then an appropriate inspection test must be carried out in accordance with the
manufacturer’s specifications.
• When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement
to these Operating Instructions.
"
Caution!
Use only original Hauser parts.
Fig. 41, installation and removal:
1.
Loosen the screws and open the hinged cover (1) of the housing.
2.
Remove the screws securing the electronics module (2). Then push up electronics module and
pull it as far as possible out of the wall-mount housing.
3.
Disconnect the plug of the sensor signal cable (7.1) including , HistoROM/S-DAT (7.2) and
HistoROM/T-DAT (7.3) from the board.
4.
Remove the cover (4) from the electronics compartment by loosening the screws.
5.
Disconnect the ribbon cable (1.2) of the display module from the amplifier board.
6.
Remove the boards (6, 7, 8):
Insert a thin pin into the hole (5) provided for the purpose and pull the board clear of its holder.
7.
Installation is the reverse of the removal procedure.