Proline t-mass 65 PROFIBUS DP
Commissioning
67
6.8
Adjustment
6.8.1
Zero point adjustment
At zero flow conditions, the output of most thermal mass flow devices has a strong dependency on
the process pressure.
The effect on the true zero point of the device, by the static line pressure, is dependant on the gas
type and the application demands and in many cases the use of the low flow cut-off function is
adequate to zero the device output. Consequently, the zero point adjustment is generally not
necessary for t-mass!
However, with some gases and /or a combination of high static line pressures, the zero point may
need to be adjusted under process conditions to restore the very low measurement capability of the
device.
Therefore, the zero point adjustment is advisable in the following special cases:
• To achieve highest measuring accuracy also with very small flow rates.
• Under process or operating conditions where the gas properties differ strongly from air properties
e.g. Hydrogen and Helium.
Preconditions for a zero point adjustment
Note the following before you perform a zero point adjustment:
• A zero point adjustment can be performed only with gases that contain no solid contents.
• The adjustment is performed with the process gas at zero flow and at operating pressure. This can
be achieved, for example, with shut-off valves upstream and/or downstream of the sensor.
– Normal operation
→
valves 1 and 2 open
– Zero point adjustment
→
valve 1 open / valve 2 closed
– Zero point adjustment
→
valve 1 closed / valve 2 open
a0003601
Fig. 37:
Zero point adjustment and shut-off valves
!
Note!
You can view the currently valid zero point value using the ZERO POINT function in the SENSOR
DATA group (see the “Description of Device Functions” manual, BA0114D/06/…).
1
2
Содержание Proline t-mass 65 PROFIBUS DP
Страница 2: ......
Страница 46: ...Operation Proline t mass 65 PROFIBUS DP 46 ...
Страница 106: ...www endress com worldwide BA113D 06 en 02 06 71021608 FM SGML6 0 ProMoDo ...