Proline t-mass 65 PROFIBUS DP
Commissioning
49
QUICK SETUP - COMMISSION
Use the
O
or
S
key at the prompt “QS-COMMISSION NO” and the device access code entry
appears. Enter the device access code “65” and press
F
; programming is enabled.
The prompt “QS-COMMISSION NO” appears. Use the
O
or
S
key to change “NO” to “YES”
and press
F
.
LANGUAGE
Use the
O
or
S
key to select the required language and continue with
F
.
PRE-SETTING
m
Select ACTUAL SETTINGS to continue programming the device and go to the next level or select
DELIVERY SETTINGS to reset the device. The device restarts and returns to the Home position.
• ACTUAL SETTINGS are the actual programmed parameters in the device.
• DELIVERY SETTINGS are the programmed parameters (factory settings plus customer specific
settings) originally delivered with the device.
SYSTEM UNITS
Select required system unit function and carry out parameterization or select QUIT to return to the
QUICK SETUP function if no further programming is required.
•
n
Only units not yet configured in the current setup are available for selection in each cycle.
•
o
The “YES” option remains visible until all the units have been configured.
“NO” is the only option displayed when no further units are available.
PIPE TYPE
p
Select the type of pipe for setting up the insertion sensor.
• Select CIRCULAR for round pipes or RECTANGULAR for duct systems.
• Use internal dimensions only.
Automatic configuration of the display
q
The “automatic parameterization of the display” option contains the following basic
settings/factory settings.
• YES: main line = MASS FLOW, additional line = TOTALIZER 1.
• NO: The existing (selected) settings remain.
The Quick Setup is now complete.
!
Note!
• The PIPE TYPE function is only available if an insertion type sensor is in use.
• For more information, see function group SENSOR DATA (see “Description of Device Functions”
manual, BA114D/06/…).
• The gas process pressure must be entered for all device types in the function PROCESS
PRESSURE, except in case of a remote pressure input being used.
For more information, see function group PROCESS PARAMETERS (see “Description of Device
Functions” manual, BA114D/06/…).