Trouble-shooting
Proline t-mass 65 PROFIBUS DP
82
9.3
Process error messages
!
Note!
Also observe the information on the following pages:
Page 40 and
9.3.1
List of process error messages
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Remedy / spare parts
Qu
a
lit
y
c
o
d
e
(H
EX
)
Me
as
ure
d
vari
a
bl
e
s
tatus
Qu
a
li
ty
st
a
tu
s
Q
u
al
it
y subs
tatus
Li
m
its
P = Process error
$
= Fault message (with
an
effect on the current operation)
! = Notice message (without
any
effect on the current operation)
422
P: FLOW LIMIT
$
: # 422
0x13
BAD
Sensor
Failure
Constant
Meas. flow
exceeded max limit
Cause:
The measured flow has exceeded the maximum
limit.
Remedy:
Reduce the flow rate or replace the instrument
with a suitable size for the application.
!
Note!
Error can be configured as a fault or notice
message.
432
P: FLUID TEMP UNST.
$
: # 432
0x13
BAD
Sensor
Failure
Constant
Gas temperature
unstable
Cause:
The gas temperature is unstable. Measurement
errors may have occured.
Remedy:
Stabilize process conditions or relocate the
instrument to a more stable measuring location.
435
P: FLOW EXT
!: # 435
0x40
0x41
0x42
UNCERTAIN
O.K.
O.K.
Low
High
Flow in extended
mode
Cause:
The flow is measured using extendend mode
(beyond calibration).
Remedy:
Resume normal operation (accuracy of device
maybe affected) or reduce the flow rate.
451
P: 0 ZERO-ADJ. N. OK.
$
: # 451
0x13
BAD
Sensor
Failure
Constant
Zero adjust is not
ok
Cause:
The saved zero point is inaccurate possibly due to
unstable process or flow conditions.
Remedy:
Stabilize process conditions or relocate the
instrument to a more stable measuring location.