Biomed functions are described in the following table:
Quick Reference – Biomed Menu
Menu Item
Description
Settings/Procedures
The
Logs
Submenu
This menu displays clinical and technical diagnostic records. Open the Biomed menu (page 2-21)
and click on
Logs
, then follow the procedures described in this table.
Component Log
The monitor maintains read-only
logs for principal components or
devices. These logs include a part
number, revision numbers, serial
numbers, software version, and
compatibility information.
1) Scroll to the component you
wish to inspect and click on it.
The log appears with the exit
arrow already selected.
2) Return to the
Logs
submenu by
clicking again.
Status Log
Displays information about your
current software and hardware
versions.
The display is read-only.
Diagnostic Log
Captures data about hardware and
software performance relating to the
monitor’s operation.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
CPS/IDS Diag. Log
Captures data about CPS and IDS
performance relating to the monitor’s
operation.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
FE Diag. Log
Captures and displays data about
front-end performance related to the
monitor’s operation.
Display is read-only.
• Click on the down arrow at the
bottom of screen to scroll
through the display.
Copy All Logs
Downloads status logs and
diagnostic logs to a memory card.
Click on
Copy All Logs
. A
confirmation message appears in
the message area to indicate the
download is complete.
Print Log
Prints an expanded version of the
Diagnostic Log to an Infinity
network
laser printer
Click on
Print Log
.
The
Service
submenu
The Service menu is password-protected
and intended for the hospital’s technical personnel or DrägerService personnel only.
Test Pulse
Test Pulse
Tests the ECG and EEG signal
clarity and display.
Clicking on
Test Pulse
causes the
following to happen:
• Injects a 300 ms pulse into the
ECG waveform (1 mV on leads I
and III, 2 mV on lead II).
• Superimposes a 100 µV pulse on
all EEG channels for 200 ms.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...