14. A zero adjustment of the pressure transducer is now necessary. The following
table outlines zeroing procedures:
15. Press the
C.O. Start
fixed key on the front of the PiCCO pod or p-CO start in
the PiCCO main selection. A p-CO averaging screen appears on the screen. If
you do not see a blood temperature baseline, exit the screen and repeat this
step.
Single Transducer Zero
Simultaneous “Smart Zero”
1. Make sure the transducer is at heart level. Dräger recommends securing the transducer holders
on the front of the PiCCO pod for proper height.
2. Close the transducer stopcock to the patient and open it to air.
3(a). Click on the parameter box associated with the
transducer you want to zero (ART, CVP, etc.). The
parameter setup menu appears.
NOTE
: You can also access the parameter menu as
follows:
1) Press the
Menu
fixed key to display the Main
menu.
2) Click on
Patient Setup.
3) Click on
Parameters.
4) Scroll to the desired pressure parameter and click.
3(b). Press the
key on the
hemodynamic pod/MPod to zero all
pressures whose transducers are open to
air.
WARNING: Do not use the ‘Smart Zero’
function if any pressure waveform is
flat (nearly static). Only use the ‘Smart
Zero’ function when all the stopcocks
are opened to air.
NOTE
: If you use this step, 3(b), and are
unable to zero a particular IBP with the
key, use the associated parameter
box as described in 3(a). This method can
be more effective.
4. Click on
Zero
.
NOTE
: If the procedure is successful, the monitor displays the message:
<IBP> zero accepted
. If
the procedure fails, the monitor displays the message:
<IBP> did not zero
.
Check the waveform. If
spikes exceed three millimeters, repeat the procedure. If procedure fails after two attempts,
replace the transducer or consult your hospital’s technical personnel.
NOTE:
Make sure all patient connections are secure and valid blood temperature and
injectate temperature readings are present before initiating a p-CO measurement.
NOTE:
In order to avoid disruption of p-CO thermodilution measurement, avoid
connecting or disconnecting any cables or changing any menus when
measurements are in progress.
During a measurement, if p-IT or p-BT readings become invalid or go out of
range, the averaging screen will disappear.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...