M
ONITORING
S
PECIFICATIONS
VF8
D
ELTA
/D
ELTA
XL/K
APPA
B-43
Pulse Oximetry
(Masimo SET SpO
2
via Pod )
Intendeduse
The Infinity Masimo SET® pod and accessories are indicated for use
during both motion and non-motion conditions and for patients who
are well or poorly perfused.
Parameter display:
Saturation (%SpO
2
), pulse rate, perfusion
Measurement range:
%SpO
2
: 1 to100 %
Pulse rate:25 to 240 /min
Perfusion:0.02 to 20 %
Measurement accuracy
(1, 5)
:
The following are for all Masimo LNCS and LNOP sensors.
Saturation (%SpO2) - During No Motion Conditions:(2)
0 to 69% not specified
70 to 100%:
Adults, Pediatrics
±2
Neonates
±3
Saturation (%SpO2) - During Motion Conditions:(3, 4)
0 to 69% not specified
70 to 100%:
Adults, Pediatrics
(3)
±3
Neonates
(4)
±3
Pulse Rate (bpm) - During No Motion Conditions:
(2)
Adults, Pediatrics, Neonates
25 - 240 bpm ±3
Pulse Rate (bpm) - During Motion Conditions:
(3, 4)
Adults, Pediatrics, Neonates
25 - 240 bpm ±5
NOTES:
1) Since pulse oximeter measurements are statistically distributed, only about two-thirds of those
measurements can be expected to fall within ±1
A
rms
of the value measured by a co-oximeter.
2) The Infinity® Masimo SET® SpO2 SmartPod™ pulse oximeter with LNOPAdt sensors has been
validated for no motion accuracy in human blood studies on healthy adult volunteers in induced
hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG
monitor. This variation equals ±1
A
rms
of the value measured by a co-oximeter.
3) The Masimo SET® pod with LNOP-Adt sensors has been validated for motion accuracy in
human blood studies on healthy adult volunteers in induced hypoxia studies while performing
rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion
between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70 -
100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals ±1
A
rms
of
the value measured by a co-oximeter.
4) The Masimo SET® pod with LNOP-Neo and Neo Pt sensors has been validated for motion and
no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia
studies while performing rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm
and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia
studies in the range of 70 - 100% SpO2 against a laboratory co-oximeter and ECG monitor. 1%
has been added to the results to account for the effects of fetal hemoglobin.
5) The pulse rate accuracy has been validated on healthy adult volunteers during induced hypoxia
studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This
variation equals ±1
A
rms
of the pulse rate value measured by the ECG monitor.
Nominal wavelength:
Red:
660 nm
IR:
905 nm
Radiant power at 50 mW
pulsed:
Min:
0.13 mW
Max:
0.79 mW
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...