MONITORING
Status Messages
Message
Condition
Suggested Action
etCO
2
< #
etCO
2
> #
The upper or lower alarm limits
exceeded by value #.
• Check the patient and treat, if necessary.
• Change the alarm limits.
• Check the equipment and replace it, if
necessary.
iCO
2
< #
iCO
2
> #
RRc < #
RRc > #
Check the ventilator for:
• Inspiratory flow
• Expiratory time
• Faulty expiratory valve
RRc Apnea
No breath is detected for a
period exceeding the RRc apnea
time set by the user.
Check the patient and treat, if necessary.
etCO
2
H/W Failure
Hardware malfunction
Disconnect then reconnect the etCO
2
module/
pod. If the message persists, return the device
to the hospital’s technical personnel and try a
new one.
etCO
2
Atm.
Press. Sensor
Failure
Hardware malfunction (etCO
2
pod).
Shift to manual mode and dial in the pressure.
If an automatic pressure required, contact the
hospital’s technical personnel.
etCO
2
Calibrate Atm.
Press.
Corrupt EEPROM
Shift to manual mode and dial in pressure. If
automatic pressure required, contact the
hospital’s technical personnel.
etCO
2
Sensor
Unplugged
The etCO
2
sensor has been
disconnected.
Disconnect then reconnect the etCO
2
sensor. If
the message persists, try another sensor.
etCO
2
Sensor
Warming Up
The C
APNOSTAT
has not yet
reached a stable temperature.
Wait for the sensor to warm up (up to three
minutes at room temperature). If the message
fails to clear, contact the hospital’s technical
personnel.
etCO
2
Sensor
Failure
The C
APNOSTAT
source current
is out of range or sensor did not
warm up within 8 minutes.
Try the sensor again. If the message persists,
try a new sensor.
etCO
2
Sensor
Too warm
The external heat source is
warming the sensor.
• Replace the sensor.
• Remove the heat source.
• If the problem persists, disconnect and
reconnect the sensor.
etCO
2
Place
Sensor on
Zero Cell
The last sensor calibration failed
or sensor is not the last sensor
calibrated on this device.
Place the sensor on the zero cell and wait for
zeroing to complete.
etCO
2
Sensor
Temp Not
Stable
The sensor temperature is
unstable following warm-up.
Wait at least three minutes for the message to
disappear. If the message persists, replace the
sensor.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...