![Dräger Infinity Delta Скачать руководство пользователя страница 361](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779361.webp)
P
RECAUTIONS
VF8
D
ELTA
/D
ELTA
XL/K
APPA
21-3
CAUTION:
Do not use the airway adapter or sensor if the monitor displays the message
“Flow: Invalid Sensor.”
Verify the mechanical/spontaneous threshold on the airway pressure (Paw)
setup menu. Improper settings of this function will cause incorrect calculations
of some respiratory mechanics parameters (for example, mechanical versus
spontaneous parameters). In pressure support ventilation modes, you may need
to adjust the threshold.
Periodically check the flow sensor and tubing for excessive moisture or
secretion build-up. Although the pod automatically attempts to clear the tubing
if necessary, moisture or secretions may remain and affect measurements.
When monitor settings for CO
2
compensation and respiratory mechanics gas
composition differ from actual gas composition in the breathing circuit, small
effects in monitored respiratory mechanics parameters may result.
Fluid that accumulates in the etCO
2
/Flow and Flow airway adapter sensors or
sensor tubing may cause the reported tidal and minute volumes to be higher
than set volumes. If reported values are significantly different than expected,
manually purge the sensor tubing. If purging does not clear the fluid, replace
the airway adapter sensor.
To avoid accidental disconnections, do not apply excessive tension to any
sensor cable or tubing.
To prevent damage to sensor windows, remove the airway adapter from the
circuit whenever aerosolized medication is being delivered.
Always position the airway adapter windows vertically to prevent patient
secretions from obscuring the adapter windows.
If you are switching adapter types (for example, adult to neonatal) for etCO
2
/
Respiratory Mechanics monitoring, recalibrate the adapter as described in
chapter 19. Sensor changes do not require recalibration.
The etCO
2
/ Respiratory Mechanics Pod should be kept at or above the level of
the patient breathing circuit to prevent ingress of fluids into the pod by gravity.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...