etCO
2
Out of
Range (High)
The CO
2
value is out of range
(high).
• Check the patient and treat,if necessary.
• Recalibrate sensor.
etCO
2
Check
Airway
Adapter/Cal
The airway adapter is dirty, not
fully seated, or out or calibration.
• Make sure the adapter is properly seated.
• Clean and calibrate the airway adapter.
etCO
2
Calibrating
Sensor
Calibrating on the zero cell in
progress.
Informational message; no action required
etCO
2
Cannot
Cal. Sensor
The calibration on the zero cell
could not be completed because
of C
APNOSTAT
temperature
instability.
• Check for any heat sources warming the
sensor and remove them.
• Wait at least three minutes for the
temperature to stabilize.
etCO
2
Adapter
Failure
The airway adapter is dirty, not
fully seated, or out or calibration.
Make sure the airway adapter is properly
seated. Clean and calibrate the airway adapter
if needed.
etCO
2
Place
Sensor On
Ref Cell
The calibration on the zero cell
completed successfully.
Place sensor on the reference cell and wait for
calibration to complete.
etCO
2
Sensor
Cal. Failed
The calibration on the zero cell
failed.
Recalibrate. If the message persists, try a new
sensor.
etCO
2
Verifying
Sensor Cal
Calibrating on reference cell in
progress.
Informational message; no action required
etCO
2
Sensor
Cal. Verified
Verification completed
successfully.
Informational message; no action required
etCO
2
Calibrating
Adapter
The airway adapter calibration
(zeroing in room air) is in
progress.
Informational message; no action required
etCO
2
Cal.
Failed,
Breaths?
Breaths were detected during
the 20 second period following
the activation of the adapter cal.
key.
Make sure the sensor is not connected to the
patient’s ventilator breathing circuit or close to a
CO
2
source. Recalibrate.
etCO
2
Cannot
Cal. Adapter
The airway adapter calibration
(zeroing in room air) could not be
completed because of
C
APNOSTAT
temperature
instability, or because the
C
APNOSTAT
was on the zero cell.
• Recalibrate holding the sensor in room air
(not on zero cell).
• Wait at least eight minutes for temperature to
stabilize and recalibrate.
• Remove any heat source warming the sensor
and recalibrate.
• Remove sensor from Zero cell, place on the
adapter, and recalibrate.
etCO
2
Adapter
Cal. Accepted
The airway adapter calibration
(zeroing in room air) is
completed.
Informational message; no action required
Message
Condition
Suggested Action
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...