21-16
D
ELTA
/D
ELTA
XL/K
APPA
VF8
21 R
ESPIRATORY
M
ECHANICS
Respiratory Mechanics Parameters
The following table lists respiratory mechanics parameters, derivations, and units of
measure.
NOTE:
Spontaneous (s) and mechanical (m) rates and volumes are calculated
separately and are designated by the letters <s> or <m> after the parameter (for
example, MVe s for expired Minute Volume, spontaneous ventilation; MVe m
for expired Minute Volume, mechanical ventilation).
Resistance and compliance are not calculated in spontaneously breathing
patients. The resistance and compliance values displayed for spontaneous
breaths are zero.
For the purposes of compensation, the temperature of the breathing circuit is
assumed to be 35 °C for inspired and expired gas; the humidity of the breathing
circuit is assumed to be 50 % for inspired gas and 100 % for expired gas; and
the default gas composition values are 21 % O
2
and 0 % CO
2
for inspiratory
and 17 % and 5 % CO
2
for expiratory (only VCO2 and TVCO2 are referenced
to STPD, or standard temperature pressure dry). Therefore, TVi will, in most
cases, be equal to or greater than TVe.
Label
Description
Derivation
Units
C20/
Cdyn
C
2
0 / Cdyn Ratio
(mechanical)
Ratio of the compliance of the last 20 % of
inspiration over dynamic lung compliance
(available only when neonatal sensor is used)
Cdyn
Lung Compliance,
Dynamic (mechanical)
Ratio of change in volume over change in pressure
during inspiration
mL/
cmH
2
O
I:E
Ratio of Inspiratory to
Expiratory Time
Ratio of inspiratory time to expiratory time, using
all the breaths and updated on a breath by breath
basis
InspT
%
Percent Inspired Time
Ratio of inspiratory time to total time of breath x
100 %
%
MAP
Mean Airway Pressure
Average airway pressure measured during entire
breath cycle
cmH
2
O
MValv
Alveolar Minute Volume Average alveolar volume per minute of effective
gas exchange
L
MVe
Expired Minute Volume Average expired volume per minute, calculated
using an eight breath moving average
L/min
Pause
Pause Pressure
(mechanical)
Airway pressure held during an inspiratory pause;
calculated only in mechanical breaths
cmH
2
O
PeCO2 Mixed Expired CO2
Volume of CO
2
in a breath divided by the eight
breath moving average of total expired volume
mmHg
PEEP
Positive End-Expiratory
Pressure
Lowest positive airway pressure at end of
expiratory period
or
average airway pressure
during last 25 % of expired flow
cmH
2
O
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...