![Dräger Infinity Delta Скачать руководство пользователя страница 306](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779306.webp)
17-16
D
ELTA
/D
ELTA
XL/K
APPA
VF8
17 P
ULSE
O
XIMETRY
(S
P
O2)
Masimo SET pod
Message
Possible Cause
Suggested Action
SpO
2
> #
SpO
2
< #
The patient’s SpO
2
is outside
the current upper or lower alarm
limits.
Check the patient and treat, if necessary.
PLS > #
PLS < #
The patient’s pulse rate is
outside the current upper or
lower alarm limits.
SpO
2
< Low limit
value - 20%
(Neonatal patient
category only)
The patient’s SpO
2
is below the
current lower alarm limit by 20%
or more.
PLS Out of Range
(Low/High)
The pulse rate is outside the
measuring range of the monitor
SpO
2
Transparent
Too much light is reaching the
sensor’s light detector. This
condition usually occurs
because the sensor is off the
finger or because too much
ambient light is detected.
• Make sure that the sensor is properly
attached to the patient’s finger.
• Remove the light source.
• Cover the sensor with opaque material.
• Make sure that no ambient light can
reach the detector.
• Contact DrägerService.
SpO
2
Unrecognized
Sensor
The monitor does not recognize
the sensor connected as valid.
• Check for a defective or unapproved
sensor.
• Replace the sensor.
• Contact the hospital’s technical
personnel or DrägerService.
SpO
2
Artifact
A persistent artifact is detected.
• Make sure the SpO
2
sensor is properly
attached to the patient, patient is not
moving, and all cables are properly
connected.
• Contact DrägerService.
SpO
2
Weak signal
Pulse amplitude is too low due
to one of the following
conditions:
• Poor perfusion (shock).
• Low body temperature.
• Check the patient and treat, if
necessary.
• Make sure the SpO
2
sensor is properly
attached to the patient and all cables
are properly connected.
• Relocate the sensor to another
extremity.
• Contact DrägerService.
SpO
2
No
Measurement
The monitor has not been able
to compute a valid measurement
within the last 30 seconds
because of unstable
measurement conditions.
• Make sure the SpO
2
sensor is properly
attached to the patient at a site free of
patient motion and all cables are
properly connected.
• Contact DrägerService.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...